pubmed:abstractText |
A modified flow cytometer has been used to detect attogram quantities of aqueous rhodamine 6G by laser-induced fluorescence analysis. A detection limit of 28 attograms (35,000 molecules) was obtained, nearly two orders of magnitude better than earlier measurements. The detection limit in concentration units was 1.4 x 10(-13) mole per liter. During the 1-second measurement period, the total volume sampled was 0.42 microliter. On average, only half a rhodamine 6G molecule was present in the 6-picoliter probed volume.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Feb
|
pubmed:issn |
0036-8075
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:day |
18
|
pubmed:volume |
219
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
845-7
|
pubmed:dateRevised |
2007-3-19
|
pubmed:meshHeading | |
pubmed:year |
1983
|
pubmed:articleTitle |
Attogram detection limit for aqueous dye samples by laser-induced fluorescence.
|
pubmed:publicationType |
Journal Article,
Research Support, U.S. Gov't, Non-P.H.S.
|