Four systems have been fabricated to measure the effective service life of respirator cartridges and canisters. The four systems employ carbon monoxide, sulfur dioxide/nitrogen dioxide, ammonia/methylamine, and chlorine/hydrogen chloride as contaminants. Each system produces test concentrations in the low percentage range (typically 0.05 to 3.0%) at flow rates of 32 and 64 liters/min and at relative humidities between 20 and 95%. Suitable anlyzers have been provided to measure the downstream concentration from 1 to 500 ppm as a function of time. Holders have been designed to accommodate the wide size variation of commercially available cartridges and canisters.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0002-8894
pubmed:author
pubmed:issnType
Print
pubmed:volume
36
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
49-56
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1975
pubmed:articleTitle
Preparation of mixtures of humidified air with CO, SO2, NO2, NH3, NH2CH3, Cl2, and HCl to measure respirator cartridge and canister service life.