pubmed:abstractText |
Tissue of commercially prepared tissue heart valves were evaluated and compared with aluminium treated, fixed porcine valve tissue in vitro (tensile strength, scanning and transmission electron microscopy) and in vivo (calcification potential after subcutaneous implantation in the rat model). Valve leaflets (n = 40) were divided into four groups according to the method of treatment: Group I (fixed in 0.652% glutaraldehyde, control), Group II (fixed and treated with aluminium), Group III (fixed and treated with Toluidine blue) and Group IV (fixed and treated with watersoluble alkyl sulphate). Tensile strength was not influenced in Group II and III (p > 0.05). Group IV indicated a significant (p < 0.05) reduction in tensile strength. Scanning electron microscopy revealed damage and loss of surface endothelium in Group III and IV respectively. Transmission electron microscopy indicated damage to underlying matricial cells in Group III and IV. Calcification potential was significantly (p < 0.001) reduced in Group II to IV. We conclude that damage ultrastructure could contribute to the reduced tensile strength in Group IV and that reduced tensile strength might have an influence on the long-term durability of tissue heart valves. Antimineralization treatment of tissue heart valves does retard calcification but is yet unable to inhibit the process completely.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Aug
|
pubmed:issn |
0021-9509
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
37
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
377-83
|
pubmed:dateRevised |
2009-11-11
|
pubmed:meshHeading |
pubmed-meshheading:8698783-Animals,
pubmed-meshheading:8698783-Bioprosthesis,
pubmed-meshheading:8698783-Calcium,
pubmed-meshheading:8698783-Heart Valve Prosthesis,
pubmed-meshheading:8698783-Male,
pubmed-meshheading:8698783-Microscopy, Electron, Scanning,
pubmed-meshheading:8698783-Microscopy, Electron, Scanning Transmission,
pubmed-meshheading:8698783-Rats,
pubmed-meshheading:8698783-Rats, Wistar,
pubmed-meshheading:8698783-Tensile Strength
|
pubmed:year |
1996
|
pubmed:articleTitle |
Laboratory evaluation of tissue heart valves. A comparative assessment.
|
pubmed:affiliation |
Department of Cardiothoracic Surgery, Faculty of Medicine, University of the Orange Free State, Bloemfontein, Republic of South Africa.
|
pubmed:publicationType |
Journal Article,
Comparative Study,
In Vitro
|