We evaluated a chemiluminescence receptor assay for vitamin B12 in serum (Magic Lite; Ciba Corning Diagnostics), in which an acridinium ester label is used with magnetic particle separation. Within- and between-batch precisions were generally acceptable, except at low analyte concentrations. The reference range determined from 104 elective preoperative patients was 120-610 pmol/L, compared with 150-590 pmol/L for our in-house radioligand-binding assay. Magic Lite discriminated between normal and abnormal results as effectively as the in-house method when local reference ranges were applied. Magic Lite demonstrated a negative bias at low analyte concentrations and was unable to detect any vitamin B12 in two B12-deficient patients. Assay accuracy--judged from analytical recovery and comparisons with the in-house method and two other radioassay kits (Quantaphase, Bio-Rad Labs., and Immophase, Ciba Corning Diagnostics)--was poor at low B12 concentrations when the manufacturer's recommended two-point calibration was used. This problem was partially corrected by using a full set of calibrators.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Apr
pubmed:issn
0009-9147
pubmed:author
pubmed:issnType
Print
pubmed:volume
40
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
537-40
pubmed:dateRevised
2011-11-17
pubmed:meshHeading
pubmed:year
1994
pubmed:articleTitle
Chemiluminescence receptor assay for measuring vitamin B12 in serum evaluated.
pubmed:affiliation
Department of Laboratory Medicine, Henderson General Hospital, Hamilton, Ontario, Canada.
pubmed:publicationType
Journal Article, Comparative Study