Total transmitted power and intensity distributions from diagnostic ultrasound systems have been measured with a radiation force balance and with a miniature hydrophone and instrumentation available in many medical centers. In assessing the accuracy of absolute intensity and power determinations from measurements of acoustic pressure with a hydrophone, ultrasonic power was computed from hydrophone measurements. This power value was compared with the power determined by a radiation force technique. On five pulse echo ultrasound systems, the ratio of the power measured with a radiation force balance to the power determined with the hydrophone varied from 0.25 to 5.5. These differences are attributed mainly to the known large variations in hydrophone response as a function of frequency, and possible time, since several other possible sources of error were evaluated and estimated to have a reasonably small net effect of 25%. The errors evaluated were those dealing primarily with measurement procedures.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0094-2405
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
6
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
404-11
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:articleTitle |
Assessment of errors in intensity measurements of pulse echo ultrasound using miniature hydrophones.
|
pubmed:publicationType |
Journal Article,
Research Support, U.S. Gov't, P.H.S.,
Research Support, U.S. Gov't, Non-P.H.S.
|