Nanometre-sized particles of ultra-high molecular weight polyethylene have been identified in the lubricants retrieved from hip simulators. Tissue samples were taken from seven failed Charnley total hip replacements, digested using strong alkali and analysed using high-resolution field emission gun-scanning electron microscopy to determine whether nanometre-sized particles of polyethylene debris were generated in vivo. A randomised method of analysis was used to quantify and characterise all the polyethylene particles isolated. We isolated nanometre-sized particles from the retrieved tissue samples. The smallest identified was 30 nm and the majority were in the 0.1 microm to 0.99 microm size range. Particles in the 1.0 microm to 9.99 microm size range represented the highest proportion of the wear volume of the tissue samples, with 35% to 98% of the total wear volume comprised of particles of this size. The number of nanometre-sized particles isolated from the tissues accounted for only a small proportion of the total wear volume. Further work is required to assess the biological response to nanometre-sized polyethylene particles.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Aug
|
pubmed:issn |
0301-620X
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
90
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1106-13
|
pubmed:dateRevised |
2010-11-10
|
pubmed:meshHeading |
pubmed-meshheading:18669972-Arthroplasty, Replacement, Hip,
pubmed-meshheading:18669972-Biocompatible Materials,
pubmed-meshheading:18669972-Hip Prosthesis,
pubmed-meshheading:18669972-Humans,
pubmed-meshheading:18669972-Lubricants,
pubmed-meshheading:18669972-Lubrication,
pubmed-meshheading:18669972-Nanoparticles,
pubmed-meshheading:18669972-Polyethylenes,
pubmed-meshheading:18669972-Prosthesis Design,
pubmed-meshheading:18669972-Prosthesis Failure,
pubmed-meshheading:18669972-Time Factors
|
pubmed:year |
2008
|
pubmed:articleTitle |
Identification of nanometre-sized ultra-high molecular weight polyethylene wear particles in samples retrieved in vivo.
|
pubmed:affiliation |
Institute of Medical and Biological Engineering (iMBE), Faculty of Biological Sciences, University of Leeds, Leeds LS2 9JT, UK.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|