pubmed:abstractText
Range sidelobe artifacts which are associated with pulse compression methods can be reduced with a new method composed of pulse elongation and deconvolution (PED). While pulse compression and PED yield similar signal-to-noise ratio (SNR) improvements, PED inherently minimizes the range sidelobe artifacts. The deconvolution is implemented as a stabilized inverse filter. With proper selection of the excitation waveform an exact inverse filter can be implemented. The excitation waveform is optimized in a minimum mean square error (MMSE) sense. An analytical expression for the power spectrum of the optimal pulse is presented and several techniques to numerically optimize the excitation pulse are shown. The effects of PED are demonstrated in computer simulations as well as ultrasonic images.
pubmed:language
eng
pubmed:journal
pubmed:status
PubMed-not-MEDLINE
pubmed:issn
0885-3010
pubmed:author
pubmed:issnType
Print
pubmed:volume
45
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
98-113
pubmed:year
1998
pubmed:articleTitle
Pulse elongation and deconvolution filtering for medical ultrasonic imaging.
pubmed:affiliation
Department of Electrical and Computer, Drexel University, Philadelphia, PA 19104, USA. bhaide@atl.com).Engineering
pubmed:publicationType
Journal Article