pubmed:abstractText
We experienced an accident related to insufficient checking of the anesthesia machine. The anesthesiologist could not ventilate manually after starting anesthesia. The cause of this case was a simple mistake of not moving the switch from ventilator to bag, but he thought severe asthma attack had occurred with the patient. About 10 minutes was lost before he recognized the real cause and the patient suffered severe hypoxia. After this event we conducted a survey of "Do you really check the anesthesia machine before anesthesia using J.S.A. checklist?" on our seven anesthesiologists. Leak test was carried out almost perfectly but other test was not. Therefore we put anesthesiologists under an obligation to use the check sheet before anesthesia and file the sheet in the medical record. The safety management of anesthesia is not only to take the safety measures but also to check the actual conditions.
pubmed:language
jpn
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
0021-4892
pubmed:author
pubmed:issnType
Print
pubmed:volume
56
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1182-5
pubmed:meshHeading
pubmed:year
2007
pubmed:articleTitle
[Actual conditions of the check system for the anesthesia machine before anesthesia. Do you really check?].
pubmed:affiliation
Department of Anesthesiology, Kin-Ikyo Chuo Hospital, Sapporo 007-8505.
pubmed:publicationType
Journal Article, English Abstract, Case Reports