pubmed:abstractText
Medicines and Healthcare products Regulatory Agency's (MHRA's) Medical Device Alert MDA/2004/048 described bias in some endocrine test results obtained on a few immunoassay platforms, particularly the Bayer Advia Centaur instrument, when using blood specimens collected into Becton Dickinson (BD) Vacutainer SSTII Advance tubes. As users of BD tubes and the Advia Centaur instrument, we addressed our concerns about the quality of the results that we had previously reported by undertaking an independent study.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0004-5632
pubmed:author
pubmed:issnType
Print
pubmed:volume
43
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
481-7
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
Comparison of Bayer Advia Centaur immunoassay results obtained on samples collected in four different Becton Dickinson Vacutainer tubes.
pubmed:affiliation
Department of Clinical Biochemistry, John Radcliffe Hospital, Oxford OX3 9DU, UK. reza.moravat@orh.nhs.uk
pubmed:publicationType