pubmed:abstractText |
A hand-held, servo-controlled tissue palpation device (SCPD) was developed to measure the pressure-pain characteristics of human tissue for disease states linked to altered pressure-pain sensitivity. The design was based on an adaptive controller using force feedback to reach and maintain a desired force in spite of movements of the operator's hand holding the device or positional changes of the subject.
|
pubmed:grant | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0021-9290
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
40
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1635-40
|
pubmed:dateRevised |
2009-11-11
|
pubmed:meshHeading | |
pubmed:year |
2007
|
pubmed:articleTitle |
Servo-controlled stepped and ramped mechanical tissue algometry.
|
pubmed:affiliation |
Dental School, University of Maryland, Baltimore, MD, USA. cstohler@umaryland.edu
|
pubmed:publicationType |
Journal Article,
Research Support, N.I.H., Extramural
|