pubmed:abstractText
The home ventilator market has grown in size and complexity. The aim of this study was to determine if common home ventilators are user-friendly for trained intensive care unit (ICU) physicians. Eleven ventilator models were tested by 13 ICU physicians without practical experience in home mechanical ventilation. Six tests were defined (start-up, unlocking, mode and setting recognition, mode change, pressure setting and alarm). For each test, the physicians were timed and their performance compared with a reference time established by a technician. The physicians also had to rate their global assessment of each machine on a visual analogue scale. The start-up test was the only test for which there was no significant difference between the physicians and the technician, except for two ventilators. The physicians were slower than the technician to unlock the ventilator and change the ventilatory mode, with some complete failures during these tests and heterogeneous results between physicians and between ventilators. Mistakes occurred in close to 50% of cases during the ventilatory mode and settings recognition test. The mean time for the most rapid of the physicians for all the tests was 58+/-53 s, compared with 15+/-9 s for the technician. In conclusion, trained intensive care unit physicians perform poorly when confronted with home mechanical ventilators without specific prior training. Therefore, it is hypothesised that the user-friendliness of home ventilators for other categories of users might be questionable.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
0903-1936
pubmed:author
pubmed:issnType
Print
pubmed:volume
27
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1236-43
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2006
pubmed:articleTitle
Evaluation of the user-friendliness of 11 home mechanical ventilators.
pubmed:affiliation
Service de Pneumologie et de Réanimation, Groupe Hospitalier Pitié-Salpêtrière, Assistance Publique-Hôpitaux de Paris, 47-83, Bd de l'Hôpital, 75651 Paris Cedex 13, France.
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't, Evaluation Studies