pubmed:abstractText
The 'Hemocue' device for rapid estimation of haemoglobin concentration was evaluated in a clinical setting. Repeatable accuracy of capillary, venous and arterial samples was examined and then compared with standard laboratory venous haemoglobin estimates using a 'Coulter JT' analyser in 42 patients. The mean values for haemoglobin (g/l) and coefficient of variation were capillary 108.2 (8.0); venous 104.9 (2.2); arterial 105.9 (2.0); and laboratory venous 104.6 (1.3). Although the mean haemoglobin values were similar, capillary samples were significantly less repeatable than venous or arterial samples (Pitman test, P < 0.001). Comparison of variance between the laboratory sample and each sampling technique demonstrated that capillary samples were significantly more variable than venous or arterial samples. Peripheral skin temperature did not influence the accuracy of capillary samples. Hemocue estimations of venous samples were found to be as accurate as laboratory estimations. The lack of repeatable accuracy of capillary estimations was sufficiently large that their use cannot be recommended in clinical practice.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0310-057X
pubmed:author
pubmed:issnType
Print
pubmed:volume
20
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
497-500
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1992
pubmed:articleTitle
A clinical evaluation of the Hemocue haemoglobinometer using capillary, venous and arterial samples.
pubmed:affiliation
Department of Anaesthesia and Intensive Care, Faculty of Medicine, Chinese University of Hong Kong.
pubmed:publicationType
Journal Article