Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
5
pubmed:dateCreated
1976-9-1
pubmed:abstractText
The performance of the Radiometer ABL 1 blood-gas analyser has been evaluated using tonometered blood samples of known PO2 and PCO2. The pH electrode was compared with two other pH electrode systems and the haemoglobin estimations were compared with measurements made by the standard cyanmethaemoglobin method. The accuracy of the machine compared well with standard blood-gas equipment, but the haemoglobin values had a small error because of a zero offset in the machine. The analyzer is simple to operate and has low running costs.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
0007-0912
pubmed:author
pubmed:issnType
Print
pubmed:volume
48
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
487-94
pubmed:dateRevised
2000-12-18
pubmed:meshHeading
pubmed:year
1976
pubmed:articleTitle
Towards blood-gas autoanalysis an evaluation of the radiometer ABL.
pubmed:publicationType