An isothermal flowmeter for the determination of local tissue blood flow is described. Flow is determined by the measurement of the thermal conductivity of the tissue in the vicinity of a heated thermistor maintained at a fixed temperature difference above a reference thermistor. Direct heating of the thermistor is utilized to eliminate the need for specially constructed indirectly heated thermistors. This design results in a device with a voltage output directly proportional to tissue thermal conductivity and to tissue blood flow. The device is shown to be adequate for the qualitative measurement of myocardial blood flow under various situations. Construction is simplified and the size of the circuit reduced by the use of readily available integrated circuits.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jul
pubmed:issn
0021-8987
pubmed:author
pubmed:issnType
Print
pubmed:volume
39
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
170-3
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1975
pubmed:articleTitle
A thermal method for the determination of tissue blood flow.