pubmed:abstractText
The high sensitivity of SF-ICP-MS (sector field inductively coupled plasma mass spectrometry) using a torch with the "guard-electrode" (capacitive decoupled plasma) allows the determination of 238U (isotope abundance 99.2%) and 235U (0.8%) and their isotope ratio in human urine samples down to the physiological level of <10 ng/l total uranium. For sample preparation UV photolysis was used. Some quality criteria like for the detection limit, the reproducibility, recovery and the isotope ratio are given. The method can be applied in occupational as well as in environmental medicine because of its outstanding detection power.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
1570-0232
pubmed:author
pubmed:copyrightInfo
Copyright 2002 Elsevier Science B.V.
pubmed:issnType
Print
pubmed:day
5
pubmed:volume
778
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
275-8
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
Determination of 235U and 238U in urine samples using sector field inductively coupled plasma mass spectrometry.
pubmed:affiliation
GSF - Research Centre for Environment and Health, Institute of Ecological Chemistry, D-85758 Neuherberg, Germany.