This study evaluated the quantitative measurements of three-dimensional (3D) volume images using multidetector row computed tomography (MDCT) and one skull specimen. Twenty-one linear distances were measured five times each by vernier caliper. A dry human skull was imaged with MDCT for various acquisition parameters at slice thicknesses of 1.25, 2.50, 3.75, and 5.00 mm for the different acquisition modes of axial and spiral scan. The distance of each corresponding item displayed on 3D rendered images was measured seven times by an uninvolved observer using a 3D software tool. Data analysis was performed to determine if there were any statistically significant differences in acquisition parameters. No significant image differences were found among the scan modes for each slice thickness. For a given scan mode, acquisition slice thickness was the important factor for quantitative measurement of 3D rendered CT images.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0897-1889
pubmed:author
pubmed:issnType
Print
pubmed:volume
15 Suppl 1
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
254-7
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
Quantitative evaluation of acquisition parameters in three-dimensional imaging with multidetector computed tomography using human skull phantom.
pubmed:affiliation
BK21 Project for Medical Sciences, Research Institute of Radiological Science, Seoul, Korea.
pubmed:publicationType
Journal Article