Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
3
pubmed:dateCreated
1977-1-25
pubmed:abstractText
An inexpensive isolated current monitor has been developed as an effort to minimize stimulus artifact in experimental systems which include iontophoresis as the source of stimulus. The monitor becomes an element of the iontophoresis circuit, but provides an output voltage isolated from the iontophoresis current. The voltage is proportional to the iontophoresis current and will represent currents of 1/2 nA to 5 muA. It is this ability to detect minute currents that is valuable in iontophoresis measurements.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0019-0578
pubmed:author
pubmed:issnType
Print
pubmed:volume
15
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
253-7
pubmed:dateRevised
2000-12-18
pubmed:meshHeading
pubmed:year
1976
pubmed:articleTitle
An isolated current monitor for iontophoresis.
pubmed:publicationType
Journal Article