pubmed:abstractText |
Validation of an open-circuit multibreath nitrogen washout technique (MBNW) for measurement of functional residual capacity (FRC). The accuracy of FRC measurement with and without continuous viscosity correction of mass spectrometer delay time (TD) relative to gas flow signal and the influence of baseline FIO2 was investigated.
|
pubmed:language |
eng
|
pubmed:journal |
|
pubmed:citationSubset |
IM
|
pubmed:chemical |
|
pubmed:status |
MEDLINE
|
pubmed:month |
May
|
pubmed:issn |
0342-4642
|
pubmed:author |
|
pubmed:issnType |
Print
|
pubmed:volume |
24
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
487-93
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:9660266-Adult,
pubmed-meshheading:9660266-Breath Tests,
pubmed-meshheading:9660266-Calibration,
pubmed-meshheading:9660266-Critical Illness,
pubmed-meshheading:9660266-Functional Residual Capacity,
pubmed-meshheading:9660266-Humans,
pubmed-meshheading:9660266-Lung,
pubmed-meshheading:9660266-Lung Diseases, Obstructive,
pubmed-meshheading:9660266-Mass Spectrometry,
pubmed-meshheading:9660266-Models, Biological,
pubmed-meshheading:9660266-Models, Structural,
pubmed-meshheading:9660266-Nitrogen,
pubmed-meshheading:9660266-Postoperative Period,
pubmed-meshheading:9660266-Reproducibility of Results,
pubmed-meshheading:9660266-Respiration, Artificial,
pubmed-meshheading:9660266-Respiratory Distress Syndrome, Adult,
pubmed-meshheading:9660266-Time Factors,
pubmed-meshheading:9660266-Viscosity
|
pubmed:year |
1998
|
pubmed:articleTitle |
Determination of functional residual capacity (FRC) by multibreath nitrogen washout in a lung model and in mechanically ventilated patients. Accuracy depends on continuous dynamic compensation for changes of gas sampling delay time.
|
pubmed:affiliation |
Zentrum Anaesthesiologie, Rettungs- u. Intensivmedizin, Göttingen, Germany. hwrigge@gwdg.de
|
pubmed:publicationType |
Journal Article
|