pubmed:abstractText
Two commercially available bacterial filters to be used as part of the mechanical ventilation unit during anaesthesia were tested for hygienic criteria. Manufacturers claim that bacterial breathing filters have a filtration capacity of about 99.995%, so that there would be no need for thermal disinfection of tubing and ventilation circuits after each use. One filter is designed for a single use only, the other can be used up to 24 times after sterilisation. Both filters consist of hydrophobic glass fibres.
pubmed:commentsCorrections
pubmed:language
ger
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0003-2417
pubmed:author
pubmed:issnType
Print
pubmed:volume
45
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
814-8
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1996
pubmed:articleTitle
[Comparative study of the efficiency of bacterial filters in long-term mechanical ventilation].
pubmed:affiliation
Klinik für Anästhesiologie, Medizinische Universität zu Lübeck.
pubmed:publicationType
Journal Article, Comparative Study, English Abstract