pubmed:abstractText |
Since its introduction in 1968, the basic algorithm for automatic recognition of ventricular ectopic depolarizations, developed at Worcester Polytechnic Institute, has been implemented and modified by a number of investigators. The modified multi-template algorithm is conceptually similar to the original algorithm except that there are eight templates stored for each patient. A new template is stored whenever the match with each of the old templates yields a correlation coefficient less than 0.79. The system reads a Holter tape at 60 x real time for heart rates of 80 beats per minute, with relatively few false positives.
|