Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:dateCreated
1993-12-13
pubmed:abstractText
This study investigated the inter-observer reproducibility of measurements of the optic nerve head as carried out with a clinical optic disc biometer. This instrument employs a modification of indirect ophthalmoscopy to enable measurement of fundus structures. Measurements were made independently by two observers on 84 eyes of 47 patients. The median inter-observer differences for each measurement were as follows: maximum disc diameter, 0.085 mm; minimum disc diameter, 0.080 mm; maximum cup diameter, 0.098 mm; minimum cup diameter, 0.078 mm; disc area, 0.225 mm2; neuroretinal rim area, 0.215 mm2. Utilising the above measurements, the instrument automatically calculates a 'rim index' to take account of variations in disc size. It then interprets the disc as normal, suspicious or glaucomatous. The optic disc biometer produced seriously conflicting interpretations in 8 discs, 7 of which had been judged by both observers to have indistinct boundaries of the disc or cup.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0950-222X
pubmed:author
pubmed:issnType
Print
pubmed:volume
7 ( Pt 3)
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
452-6
pubmed:dateRevised
2009-11-3
pubmed:meshHeading
pubmed:year
1993
pubmed:articleTitle
Inter-observer variation in clinical optic disc biometry.
pubmed:affiliation
Department of Ophthalmology, Southern General Hospital, Glasgow, UK.
pubmed:publicationType