pubmed:abstractText |
This study investigated the inter-observer reproducibility of measurements of the optic nerve head as carried out with a clinical optic disc biometer. This instrument employs a modification of indirect ophthalmoscopy to enable measurement of fundus structures. Measurements were made independently by two observers on 84 eyes of 47 patients. The median inter-observer differences for each measurement were as follows: maximum disc diameter, 0.085 mm; minimum disc diameter, 0.080 mm; maximum cup diameter, 0.098 mm; minimum cup diameter, 0.078 mm; disc area, 0.225 mm2; neuroretinal rim area, 0.215 mm2. Utilising the above measurements, the instrument automatically calculates a 'rim index' to take account of variations in disc size. It then interprets the disc as normal, suspicious or glaucomatous. The optic disc biometer produced seriously conflicting interpretations in 8 discs, 7 of which had been judged by both observers to have indistinct boundaries of the disc or cup.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:issn |
0950-222X
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
7 ( Pt 3)
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
452-6
|
pubmed:dateRevised |
2009-11-3
|
pubmed:meshHeading | |
pubmed:year |
1993
|
pubmed:articleTitle |
Inter-observer variation in clinical optic disc biometry.
|
pubmed:affiliation |
Department of Ophthalmology, Southern General Hospital, Glasgow, UK.
|
pubmed:publicationType |
Journal Article
|