The aim of this study was to evaluate sensor noise with the use of the subtraction method in radiographs captured with three direct digital intraoral systems. Ten radiographs were taken of the lower left molar region of a phantom head at each of three exposure times: 0.20 seconds, 0.46 seconds, and 0.60 seconds with the use of the RadioVisioGraphy (Trophy Radiologie, Vincennes, France), Sens-a-Ray (Regam Medical Systems, AB, Sundsvall, Sweden), and Visualix (Gendex, Philips Medical Systems, Inc., Monza, Italy) systems. Neither the x-ray tube nor the phantom were moved between exposures, and the three sensors were identically positioned. The images were stored in the tagged image file format provided by the systems in 8-bit depth and imported by a subtraction program. Subtractions were performed between identical images taken with the three systems. The standard deviation for the distribution of the shades of grey in the subtraction image histogram served as an expression for image noise. Paired t tests evaluated differences between the standard deviations of the subtraction images from the three systems. The standard deviation increased with increasing exposure time for all three systems (p < 0.00001). The standard deviation for the images performed with Visualix were 6.47, 10.34, and 11.16 at exposure times 0.20, 0.46, and 0.60, respectively. For the RadioVisioGraphy, these values were 1.61, 2.03, and 2.18, and for Sens-a-Ray 2.90, 3.98, and 3.96, respectively. The differences between the systems were highly statistically significant (p < 0.0001).(ABSTRACT TRUNCATED AT 250 WORDS)
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
D
|
pubmed:status |
MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0030-4220
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
77
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
70-4
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading |
pubmed-meshheading:8108102-Artifacts,
pubmed-meshheading:8108102-Evaluation Studies as Topic,
pubmed-meshheading:8108102-Humans,
pubmed-meshheading:8108102-Models, Structural,
pubmed-meshheading:8108102-Radiography, Dental,
pubmed-meshheading:8108102-Reproducibility of Results,
pubmed-meshheading:8108102-Semiconductors,
pubmed-meshheading:8108102-Subtraction Technique
|
pubmed:year |
1994
|
pubmed:articleTitle |
Sensor noise in direct digital imaging (the RadioVisioGraphy, Sens-a-Ray, and Visualix/Vixa systems) evaluated by subtraction radiography.
|
pubmed:affiliation |
Department of Oral Radiology, Royal Dental College, University of Aarhus, Denmark.
|
pubmed:publicationType |
Journal Article,
Comparative Study
|