pubmed:abstractText |
The effectiveness of an infrainguinal bypass graft surveillance programme using duplex scanning was assessed over a 2-year period. Of 220 infrainguinal bypass grafts (123 vein and 97 PTFE grafts; 114 to the above knee level, 94 below knee and 12 distal to popliteal artery) in 203 patients, 208 (94.5%) were available for follow-up surveillance. The protocol called for duplex scans at 6 weeks, 3, 6, 9 and 12 months. Fifty-seven grafts (27%) were found to have a V1/V2 ratio equal to or greater than 1.5 and in this group 25 grafts occluded. The median time between primary operation and positive duplex finding was 4 months. Thirty-nine grafts failed during follow-up (at time of analysis median follow-up was 12 months [range 2-83 months]). There were 18 interventions resulting from surveillance-detected stenoses. The median time between positive duplex finding and further investigation was 2 months. Further, there were significant differences in the site of abnormal findings between ePTFE and vein grafts. The value of a surveillance programme may be reduced if there are low rates of intervention and/or excessive delays in intervention following the demonstration of graft-related stenoses. Surveillance programmes and subsequent interventions need to be audited.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Oct
|
pubmed:issn |
0035-8835
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
39
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
297-300
|
pubmed:dateRevised |
2004-11-17
|
pubmed:meshHeading |
pubmed-meshheading:7861339-Adult,
pubmed-meshheading:7861339-Aged,
pubmed-meshheading:7861339-Aged, 80 and over,
pubmed-meshheading:7861339-Blood Vessel Prosthesis,
pubmed-meshheading:7861339-Female,
pubmed-meshheading:7861339-Graft Occlusion, Vascular,
pubmed-meshheading:7861339-Groin,
pubmed-meshheading:7861339-Humans,
pubmed-meshheading:7861339-Leg,
pubmed-meshheading:7861339-Male,
pubmed-meshheading:7861339-Middle Aged,
pubmed-meshheading:7861339-Polytetrafluoroethylene,
pubmed-meshheading:7861339-Retrospective Studies,
pubmed-meshheading:7861339-Veins
|
pubmed:year |
1994
|
pubmed:articleTitle |
Duplex ultrasound surveillance of infrainguinal bypass grafts: auditing the process.
|
pubmed:affiliation |
Department of Radiology, Royal Infirmary, Edinburgh, UK.
|
pubmed:publicationType |
Journal Article
|