Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1980-7-12
pubmed:abstractText
From general physical laws simple formulae have been deduced from which the rigidity of an external fixation unit mounted in different ways becomes mathematically calculable. Furthermore, the stability of various mounting systems under bending forces was investigated experimentally. Particular attention was paid to details of operating technique which are of special importance to the stability of the whole construction. The calculated data are consistent with the mechanical test results. The best arrangement of the Steinmann pins and Schanz screws is discussed, and how to mount the frame optimally, depending on the specific type and location of the fracture. Special instructions concerning the mounting of the external fixation unit are given which, depending on the specific circumstances, guarantee optimum stability of the unit.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:issn
0344-8444
pubmed:author
pubmed:issnType
Print
pubmed:volume
96
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
35-43
pubmed:dateRevised
2008-2-21
pubmed:meshHeading
pubmed:year
1980
pubmed:articleTitle
How to increase the stability of external fixation units. Mechanical tests and theoretical studies.
pubmed:publicationType
Journal Article