Two hundred and forty blood glucose readings were made on 5 Glucochek machines by 2 different operators. The results on all machines correlated poorly with equivalent results from an automatic glucose analyser with major discrepancies at all glucose concentrations. There was a wide gap in performance between the 2 operators. Analysis of variance suggested that errors were not only due to inter-operator and inter-machine differences. Possible sources of error were shown to include time of reaction and the amount of blood on Dextrostix. Dextrostix colour changes measured by eye were as accurate as those by Glucochek without the rogue values found at higher glucose concentrations with Glucochek.
pubmed:commentsCorrections
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0032-5473
pubmed:author
pubmed:issnType
Print
pubmed:volume
56
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
757-60
pubmed:dateRevised
2009-11-18
pubmed:meshHeading
pubmed:year
1980
pubmed:articleTitle
Problems in home monitoring of blood glucose.