pubmed-article:7058893pubmed:abstractTextA modification of the coulometric acid-base titration method is described for the measurement of titratable acid and ammonium in nanoliter samples. The main components consist of a pH-seeking device and a miniaturized antimony electrode system capable of delivering OH- ions at a known rate. The modifications include use of a cup-shaped pH-sensitive glass-membrane electrode for monitoring pH changes during titration and a new electronic titration circuit consisting of a high-impedance differential electrometer, a constant-current source, and an electronic integrating circuit. Details regarding the construction of the components of the titration system and practical aspects of its use are also provided.lld:pubmed
pubmed-article:7058893pubmed:granthttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:7058893pubmed:languageenglld:pubmed
pubmed-article:7058893pubmed:journalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:7058893pubmed:citationSubsetIMlld:pubmed
pubmed-article:7058893pubmed:chemicalhttp://linkedlifedata.com/r...lld:pubmed
pubmed-article:7058893pubmed:statusMEDLINElld:pubmed
pubmed-article:7058893pubmed:monthJanlld:pubmed
pubmed-article:7058893pubmed:issn0002-9513lld:pubmed
pubmed-article:7058893pubmed:authorpubmed-author:GiebischGGlld:pubmed
pubmed-article:7058893pubmed:authorpubmed-author:JaegerPPlld:pubmed
pubmed-article:7058893pubmed:authorpubmed-author:KarlmarkBBlld:pubmed
pubmed-article:7058893pubmed:authorpubmed-author:FeirDDlld:pubmed
pubmed-article:7058893pubmed:issnTypePrintlld:pubmed
pubmed-article:7058893pubmed:volume242lld:pubmed
pubmed-article:7058893pubmed:ownerNLMlld:pubmed
pubmed-article:7058893pubmed:authorsCompleteYlld:pubmed
pubmed-article:7058893pubmed:paginationF95-9lld:pubmed
pubmed-article:7058893pubmed:dateRevised2007-11-14lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:meshHeadingpubmed-meshheading:7058893-...lld:pubmed
pubmed-article:7058893pubmed:year1982lld:pubmed
pubmed-article:7058893pubmed:articleTitleCoulometric acid-base titration in nanoliter samples with glass and antimony electrodes.lld:pubmed