1982-2-25
|
pubmed:abstractText |
In 43 percent of 30 consecutive open heart surgery patients, Swan-Ganz catheter tips lodged within 1 cm of or above the left atrium. When in this position the wedge pressure measured by the catheter was not an accurate estimate of left atrial pressure when positive end-expiratory pressure (PEEP) was used, especially when left atrial pressure was low. Catheters located below the left atrium were accurate at all levels of PEEP tested. The position of Swan-Ganz catheters should be confirmed by a lateral chest roentgenogram when PEEP is used, and catheter tips not below the left atrium should be repositioned.
|
pubmed:grant | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
AIM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Dec
|
pubmed:issn |
0012-3692
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
80
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
666-70
|
pubmed:dateRevised |
2007-11-14
|
pubmed:meshHeading |
pubmed-meshheading:7030651-Atrial Function,
pubmed-meshheading:7030651-Blood Pressure,
pubmed-meshheading:7030651-Cardiac Surgical Procedures,
pubmed-meshheading:7030651-Heart Catheterization,
pubmed-meshheading:7030651-Humans,
pubmed-meshheading:7030651-Intermittent Positive-Pressure Ventilation,
pubmed-meshheading:7030651-Positive-Pressure Respiration,
pubmed-meshheading:7030651-Pulmonary Wedge Pressure,
pubmed-meshheading:7030651-Radiography, Thoracic
|
pubmed:year |
1981
|
pubmed:articleTitle |
Swan-Ganz catheter location and left atrial pressure determine the accuracy of the wedge pressure when positive end-expiratory pressure is used.
|
pubmed:publicationType |
Journal Article,
Research Support, U.S. Gov't, P.H.S.
|