pubmed:abstractText |
A highly sensitive and simple fluorimetric method for the determination of free and total polyamines, spermidine, spermine, putrescine and cadaverine, in human serum by high-performance liquid chromatography is described. The polyamines, obtained after clean-up of deproteinized serum by Cellex P column chromatography, are converted to their fluorescamine derivatives in the presence of nickel ion which inhibits the reaction of interfering amines with fluorescamine, and the derivatives are separated simultaneously by reversed-phase chromatography (LiChrosorb RP-18) with a linear gradient elution. The lower limits of detection are 10 and 15 pmole for spermine and the others in 0.5 ml of serum, respectively.
|