pubmed:abstractText
A microporous-coated femoral prosthesis with an average pore size of 80 to 100 microns was retrieved after seven years in a revision operation on a failed cemented acetabular component. The proximal anterior and lateral aspects of the implant were encrusted with bone. The bone was attached to the porous coating in regions of ingrowth. The implant was stable within the femur although most of the porous coating was not bone ingrown. There was progressive bone resorption in the proximal femur, suggestive of a reactive biologic mechanism. There was also a sizable posterior pocket of granulation tissue containing chronic inflammatory round cells associated with numerous birefringent polymeric particles. Polyethylene wear debris and acrylic cement were engulfed by multinucleated giant cells enveloping the loose cup.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
AIM
pubmed:status
MEDLINE
pubmed:month
Nov
pubmed:issn
0009-921X
pubmed:author
pubmed:issnType
Print
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
303-10
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
1987
pubmed:articleTitle
Histologic analysis of a retrieved microporous-coated femoral prosthesis. A seven-year case report.
pubmed:affiliation
Anderson Orthopaedic Research Institute, Arlington, Virginia.
pubmed:publicationType
Journal Article, Case Reports, Research Support, Non-U.S. Gov't