A modified procedure for measuring cyclosporin in whole blood by high-performance liquid chromatography is described and evaluated for clinical use. Sample preparation uses solid-phase extraction cartridges that can be reused. Life of the reverse-phase analytical column exceeds 1,000 injections at 70 degrees C. Cyclosporins A and D (internal standard) elute after 5.6 and 7.6 min, respectively. Calibration plots are linear from 50 ng/ml to at least 2,000 ng/ml. Within-day and between-day imprecision is less than 9% (coefficient of variation). Minimum measurable concentration is 50 ng/ml.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
0163-4356
pubmed:author
pubmed:issnType
Print
pubmed:volume
10
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
97-100
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1988
pubmed:articleTitle
A modified assay for cyclosporin in blood using solid-phase extraction with high-performance liquid chromatography.
pubmed:affiliation
Department of Clinical Pharmacology, Princess Alexandra Hospital, Brisbane, Queensland, Australia.
pubmed:publicationType
Journal Article, Clinical Trial