A modified procedure for measuring cyclosporin in whole blood by high-performance liquid chromatography is described and evaluated for clinical use. Sample preparation uses solid-phase extraction cartridges that can be reused. Life of the reverse-phase analytical column exceeds 1,000 injections at 70 degrees C. Cyclosporins A and D (internal standard) elute after 5.6 and 7.6 min, respectively. Calibration plots are linear from 50 ng/ml to at least 2,000 ng/ml. Within-day and between-day imprecision is less than 9% (coefficient of variation). Minimum measurable concentration is 50 ng/ml.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:issn |
0163-4356
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
10
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
97-100
|
pubmed:dateRevised |
2007-11-15
|
pubmed:meshHeading | |
pubmed:year |
1988
|
pubmed:articleTitle |
A modified assay for cyclosporin in blood using solid-phase extraction with high-performance liquid chromatography.
|
pubmed:affiliation |
Department of Clinical Pharmacology, Princess Alexandra Hospital, Brisbane, Queensland, Australia.
|
pubmed:publicationType |
Journal Article,
Clinical Trial
|