Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1 Pt 2
pubmed:dateCreated
1989-4-14
pubmed:abstractText
Within the last years, a variety of different energy sources has been investigated to test their feasibility for catheter ablation of myocardial tissue. This report summarizes our experience of the use of radiofrequency alternating current (500 kHz, unipolar mode) for coagulation of ventricular myocardium in canine experiments. Under standardized in vitro conditions, we found a significant correlation between actually delivered radiofrequency energy and assessed myocardial necrosis (r = 0.87). However, this did not hold for percutaneous application of radiofrequency alternating current to the beating dog heart (r = 0.32). In the intact dog heart, the size of induced lesions paralleled catheter contact pressure which was varied until ST-egment elevation was either 0-2 mV or 5-7 mV. However, no statistical significant differences in either calculated energy or tissue impedance were observed. Under in vivo conditions, a significant improvement in the predictability of the resulting size of lesions was observed when catheter tip temperature, measured via a built-in Ni/CrNi thermoelement, was monitored (r = 0.07). Changes in tip temperature during coagulation also indicated the quality of catheter contact, catheter damage and the appearance of carbonization at the tip of the ablation catheter. Total perforation of the myocardial wall and proarrhythmogenic effects were only rarely observed. In conclusion, catheter coagulation of myocardial tissue using radiofrequency energy can be considered as safe and effective. Since changes in catheter tip temperature occurring during coagulation were found to predict the extent of induced tissue necrosis, the development of temperature controlled radiofrequency devices seems promising and necessary.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0147-8389
pubmed:author
pubmed:issnType
Print
pubmed:volume
12
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
187-95
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1989
pubmed:articleTitle
Coagulation of ventricular myocardium using radiofrequency alternating current: bio-physical aspects and experimental findings.
pubmed:affiliation
Department of Cardiology-Angiology, University Hospital, Münster, Germany, FR.
pubmed:publicationType
Journal Article, Review