pubmed:abstractText
The safety of a new portable pump (Baxter Infusor), was evaluated for continuous epidural infusion of local anesthetics. The experimental flows at different temperatures were 1.87 ml.hr-1 at 28 degrees C and 2.14-2.15 ml.hr-1 at 33 degrees C. The average temperature of local anesthetics in this pump attached to patients was 29 degrees C. Concentration of mepivacaine reached the maximal value (1.129 +/- 0.439 micrograms.ml-1) at 12 hours after the application of the infusor. The concentration of mepivacaine was maintained at a constant level from 6th hour to 24th hour. The result indicates that Infusor is a safe device for continuous infusion of local anesthetics because of constant flow and low level of local anesthetics.
pubmed:language
jpn
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Sep
pubmed:issn
0021-4892
pubmed:author
pubmed:issnType
Print
pubmed:volume
39
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1245-51
pubmed:dateRevised
2007-11-15
pubmed:meshHeading
pubmed:year
1990
pubmed:articleTitle
[Evaluation of the safety of a new portable pump for continuous epidural infusion of local anesthetics].
pubmed:affiliation
Department of Anesthesiology, Hiraka General Hospital, Yokote.
pubmed:publicationType
Journal Article, English Abstract