pubmed:abstractText |
Beam-shaping or "bow tie" (BT) filters are used to spatially modulate the x-ray beam in a CT scanner, but the conventional method of step-and-shoot measurement to characterize a beam's profile is tedious and time-consuming. The theory for characterization of bow tie relative attenuation (COBRA) method, which relies on a real-time dosimeter to address the issues of conventional measurement techniques, was previously demonstrated using computer simulations. In this study the feasibility of the COBRA theory is further validated experimentally through the employment of a prototype real-time radiation meter and a known BT filter.
|
pubmed:grant | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Mar
|
pubmed:issn |
0094-2405
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
38
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
1406-15
|
pubmed:meshHeading | |
pubmed:year |
2011
|
pubmed:articleTitle |
Experimental validation of a method characterizing bow tie filters in CT scanners using a real-time dose probe.
|
pubmed:affiliation |
Department of Radiology, University of California, Davis Medical Center, Rm. 0505, ACC Ellison Bldg., 4860 Y St., Sacramento, California 95817, USA.
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't,
Research Support, N.I.H., Extramural,
Validation Studies
|