A soft contact-lens biosensor (SCL-biosensor) for novel non-invasive biomonitoring of tear fluids was fabricated and tested. Wearing a biosensor on eye enabled the in situ monitoring of tear contents. The biosensor has an enzyme immobilized electrode on the surface of a polydimethyl siloxane (PDMS) contact lens. The SCL-biosensor was fabricated using microfabrication techniques for functional polymers (PDMS and 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer). In investigation of in vitro characterization, the SCL-biosensor showed excellent relationship between the output current and glucose concentration from 0.03 to 5.0 mmol·L(-1), with a correlation coefficient of 0.994. The calibration range covered the reported tear glucose concentrations (0.14 mmol·L(-1)). Based on the result, ocular biomonitoring with the SCL-biosensor was carried out. The SCL-biosensor well worked both in the static state and the dynamic state. The tear glucose level of rabbit was estimated to 0.12 mmol·L(-1) at first and then the tear turnover was successfully calculated to be 29.6?±?8.42% min(-1). The result indicated that SCL-biosensor is useful for advanced biomonitoring on eye.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Aug
pubmed:issn
1572-8781
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
13
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
603-11
pubmed:meshHeading
pubmed:year
2011
pubmed:articleTitle
Biomedical soft contact-lens sensor for in situ ocular biomonitoring of tear contents.
pubmed:affiliation
Department of Advanced Science and Technology for Biomedical Sensors, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8549, Japan.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't