Source:http://linkedlifedata.com/resource/pubmed/id/19758673
Switch to
Predicate | Object |
---|---|
rdf:type | |
lifeskim:mentions | |
pubmed:issue |
2
|
pubmed:dateCreated |
2009-11-17
|
pubmed:abstractText |
Medical imaging is a vital component of high intensity focused ultrasound (HIFU) therapy, which is gaining clinical acceptance for tissue ablation and cancer therapy. Imaging is necessary to plan and guide the application of therapeutic ultrasound, and to monitor the effects it induces in tissue. Because they can transmit high intensity continuous wave ultrasound for treatment and pulsed ultrasound for imaging, dual-mode transducers aim to improve the guidance and monitoring stages. Their primary advantage is implicit registration between the imaging and treatment axes, and so they can help ensure before treatment that the therapeutic beam is correctly aligned with the planned treatment volume. During treatment, imaging signals can be processed in real-time to assess acoustic properties of the tissue that are related to thermal ablation. Piezocomposite materials are favorable for dual-mode transducers because of their improved bandwidth, which in turn improves imaging performance while maintaining high efficiency for treatment. Here we present our experiences with three dual-mode transducers for interstitial applications. The first was an 11-MHz monoelement designed for use in the bile duct. It had a 25x7.5 mm(2) aperture that was cylindrically focused to 10mm. The applicator motion was step-wise rotational for imaging and therapy over a 360 degrees, or smaller, sector. The second transducer had 5-elements, each measuring 3.0x3.8 mm(2) for a total aperture of 3.0x20 mm(2). It operated at 5.6 MHz, was cylindrically focused to 14 mm, and was integrated with a servo-controlled oscillating probe designed for sector imaging and directive therapy in the liver. The last transducer was a 5-MHz, 64-element linear array designed for beam-formed imaging and therapy. The aperture was 3.0x18 mm(2) with a pitch of 0.280 mm. Characterization results included conversion efficiencies above 50%, pulse-echo bandwidths above 50%, surface intensities up to 30 W/cm(2), and axial imaging resolutions to 0.2 mm. The second transducer was evaluated in vivo using porcine liver, where coagulation necrosis was induced up to a depth of 20 mm in 120 s. B-mode and M-mode images displayed a hypoechoic region that agreed well with lesion depth observed by gross histology. These feasibility studies demonstrate that the dual-mode transducers had imaging performance that was sufficient to aid the guidance and monitoring of treatment, and could sustain high intensities to induce coagulation necrosis in vivo.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Feb
|
pubmed:issn |
1874-9968
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:volume |
50
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
216-20
|
pubmed:meshHeading |
pubmed-meshheading:19758673-Acoustics,
pubmed-meshheading:19758673-Animals,
pubmed-meshheading:19758673-Bile Ducts,
pubmed-meshheading:19758673-Electronics, Medical,
pubmed-meshheading:19758673-Equipment Design,
pubmed-meshheading:19758673-Feasibility Studies,
pubmed-meshheading:19758673-Image Processing, Computer-Assisted,
pubmed-meshheading:19758673-Liver,
pubmed-meshheading:19758673-Models, Theoretical,
pubmed-meshheading:19758673-Swine,
pubmed-meshheading:19758673-Transducers,
pubmed-meshheading:19758673-Ultrasonic Therapy,
pubmed-meshheading:19758673-Ultrasonography, Interventional
|
pubmed:year |
2010
|
pubmed:articleTitle |
Dual-mode transducers for ultrasound imaging and thermal therapy.
|
pubmed:affiliation |
Inserm, U556, Lyon F-69003, France. nowen@apl.washington.edu
|
pubmed:publicationType |
Journal Article,
Research Support, Non-U.S. Gov't
|