pubmed:abstractText |
This study was designed to discriminate, according to their accuracy, between three ambulatory pressurometers (Diasys 200R, Novacor; P IV, Del Mar Avionics; SpaceLab 90202, SpaceLab). The evaluation was performed against invasive arterial reference measurements. Accuracy was assessed by calculating the error on pressure (EOP) as the difference between invasive and non-invasive measurement of arterial blood pressure. For the systolic values, accuracy (mean of EOP differences) and uncertainty (SD of these differences) were -0.9 +/- 9.7, -4.3 +/- 10.1 and -16.7 +/- 10.1 mmHg for, respectively, Diasys, PIV and SpaceLab. For diastolic values, they were, respectively, 5.9 +/- 6.7, 6.8 +/- 8.5 and 9.1 +/- 6.6 mmHg. EOP was then separated in two different types of errors: (i) the error of dispersion appreciated by the index of homogeneity calculated by a Lehmann analysis and leading to a statistical classification (ii) the error due to the drift of EOP with the reference value, this last error being easier to correct. Two different behaviours were observed for the EOP: (i) the drift of EOP of systolic values was significantly larger for the oscillometric (SpaceLab) than for the auscultatory (Diasys and P IV) method, with no difference between Diasys and P IV (ii) the homogeneity index was not statistically different among these three devices. These data suggest that, in case the correction of the drift of EOP is carried out, there is no statistical significant difference in accuracy between these three pressurometers. However, in our experimental conditions, the two ambulatory pressurometers recording the Korotkoff sounds have a better accuracy than the one using the oscillometric approach.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Sep
|
pubmed:issn |
0036-5513
|
pubmed:author | |
pubmed:issnType |
Print
|
pubmed:volume |
51
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
461-6
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading |
pubmed-meshheading:1947731-Adult,
pubmed-meshheading:1947731-Aged,
pubmed-meshheading:1947731-Blood Pressure Determination,
pubmed-meshheading:1947731-Blood Pressure Monitors,
pubmed-meshheading:1947731-Female,
pubmed-meshheading:1947731-Humans,
pubmed-meshheading:1947731-Male,
pubmed-meshheading:1947731-Middle Aged
|
pubmed:year |
1991
|
pubmed:articleTitle |
Accuracy of ambulatory blood pressure determination: a comparative study.
|
pubmed:affiliation |
Laboratoire de Physiologie-GIP Exercise, CHU Saint-Etienne, Saint Jean Bonnefonds, France.
|
pubmed:publicationType |
Journal Article,
Comparative Study
|