pubmed:abstractText
An acetabular cup made of a compliant polycarbonate-urethane material has been introduced as an alternative bearing solution. This case study examines the use of this acetabular cup in a single patient at 12 months. The cup had minimal loss of thickness with the most being in the superior area (approximately 10%) and minimal loss of weight (2.4%). On the back side was evidence of abrasive macroscopic wear on one side of the implanted component in the area of directional loading from the head to the acetabulum. On the front side, the wear rate was determined to be minimal (less than 15 mm3 per year), meaning that most of the change in thickness and weight was caused on the back side. The retrieved synovial fluid appeared normal in color and volume at the time of revision. The histology of the tissue taken showed minimal wear particles and minimal reactivity, confirming that the patient did not have any signs of synovitis.The analysis of the cup confirms the preclinically determined low wear articulation and biocompatibility of polycarbonate-urethane as a weight-bearing material. In summary, the surgical findings, data reviewed, and images taken from this case report warrant further study.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
T
pubmed:chemical
pubmed:status
MEDLINE
pubmed:issn
1050-6934
pubmed:author
pubmed:issnType
Print
pubmed:volume
18
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
69-74
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
Retrieval analysis of a polycarbonate-urethane acetabular cup: a case report.
pubmed:affiliation
Orthopädische Klinik Kassel gGmbH, Kassel, Germany. wsiebert@okkassel.de
pubmed:publicationType
Journal Article, Case Reports