pubmed:abstractText
Ventricular safety pacing (VSP) is an algorithm used to prevent crosstalk inhibition and ventricular capture during the vulnerable period. We report a 78-year-old man with implantable dual-chamber defibrillator, in whom clusters of ventricular tachycardias (VTs) were provoked by the VSP. During rapid DDDR pacing, the delivery of the VSP after every other atrial-paced beat resulted in short-long-short ventricular sequences and induced VTs. An atrial-based lower rate timing, long atrioventricular pacing interval, and automatic gain control also accounted for this arrhythmogenic ventricular sequence. The VSP and the subsequent VT were eliminated by decreasing the pacing rate.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
1532-2092
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
10
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1238-42
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
Short-long-short sequence caused by ventricular safety pacing inducing ventricular tachycardia in a patient with a dual-chamber implantable cardioverter defibrillator.