pubmed:abstractText
We report two patients with cardiac resynchronization therapy (CRT) devices and evidence of refractory heart failure in whom impaired intraatrial conduction in one patient, and interatrial conduction in the other, prohibited optimization of the atrioventricular (AV) timing sequence. The patient with intraatrial conduction delay exhibited late right atrial sensing and latency during right atrial pacing that required programming of a short-sensed AV delay and long-paced AV delay (wide differential AV delay). In both patients AV junctional ablation and echocardiography-guided device optimization significantly improved heart failure.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
1540-8159
pubmed:author
pubmed:issnType
Electronic
pubmed:volume
31
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
685-90
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
AV junctional ablation allowing more effective delivery of cardiac resynchronization therapy in patients with intra- and interatrial conduction delay.
pubmed:affiliation
Arrhythmia Service and Division of Cardiology, Tampa General Hospital, Tampa, Florida, USA. bherweg@health.usf.edu
pubmed:publicationType
Journal Article, Case Reports