Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
2
pubmed:dateCreated
2008-5-20
pubmed:abstractText
This work describes an assay system that has been developed to quantify raltegravir concentrations in human plasma using a liquid-liquid extraction technique paired with HPLC separation and MS-MS detection. The dynamic range of this assay extends from 1 to 3000 ng/mL, with a coefficient of determination (r(2), mean+/-SD) of 0.9992+/-0.0002. The mean precision values for calibration standards ranged from 0.6% to 3.0%, while accuracy values were 96.5-104.3%. This procedure is an accurate, precise, and sensitive method for raltegravir quantitation and was successfully validated using external proficiency testing.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
May
pubmed:issn
1570-0232
pubmed:author
pubmed:issnType
Print
pubmed:day
15
pubmed:volume
867
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
165-71
pubmed:meshHeading
pubmed:year
2008
pubmed:articleTitle
A sensitive HPLC-MS-MS method for the determination of raltegravir in human plasma.
pubmed:affiliation
Division of Clinical Pharmacology, University of Alabama at Birmingham School of Medicine, Birmingham, AL 35294-0019, United States.
pubmed:publicationType
Journal Article, Research Support, N.I.H., Extramural, Validation Studies