Variation in computed tomography (CT) image gray-scale and spatial geometry due to specimen orientation, magnification, voxel size, differences in X-ray photon energy and limited field-of-view during the scan, were evaluated in repeated micro-CT scans of iliac crest biopsies and test phantoms. Using the micro-CT scanner on beamline X2B at the Brookhaven National Laboratory's National Synchrotron Light Source, 3-D micro-CT images were generated. They consisted of up to 1024 x 2400(2), 4-microm cubic voxels, each with 16-bit gray-scale. We also reconstructed the images at 16-, 32-, and 48-microm voxel resolution. Scan data were reconstructed from the complete profiles using filtered back-projection and from truncated profiles using profile-extension and with a Local reconstruction algorithm. Three biopsies and one bone-like test phantom were each rescanned at three different times at annual intervals. For the full-data-set reconstructions, the reproducibility of the estimates of mineral content of bone at mean bone opacity value, was +/-28.8, i.e., 2.56%, in a 4-microm cubic voxel at the 95% confidence level. The reproducibility decreased with increased voxel size. The interscan difference in imaged bone volume ranged from 0.86 4-microm 0.64% at 4-microm voxel resolution, and 2.64 4-microm 2.48% at 48 microm.
|
pubmed:grant | |
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:status |
MEDLINE
|
pubmed:month |
Apr
|
pubmed:issn |
1558-0062
|
pubmed:author | |
pubmed:issnType |
Electronic
|
pubmed:volume |
27
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
569-75
|
pubmed:meshHeading |
pubmed-meshheading:18390353-Algorithms,
pubmed-meshheading:18390353-Artificial Intelligence,
pubmed-meshheading:18390353-Humans,
pubmed-meshheading:18390353-Ilium,
pubmed-meshheading:18390353-Imaging, Three-Dimensional,
pubmed-meshheading:18390353-Radiographic Image Enhancement,
pubmed-meshheading:18390353-Radiographic Image Interpretation, Computer-Assisted,
pubmed-meshheading:18390353-Reproducibility of Results,
pubmed-meshheading:18390353-Sensitivity and Specificity,
pubmed-meshheading:18390353-Tomography, X-Ray Computed
|
pubmed:year |
2008
|
pubmed:articleTitle |
Reproducibility of global and local reconstruction of three-dimensional micro-computed tomography of iliac crest biopsies.
|