Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
6
pubmed:dateCreated
2007-5-28
pubmed:abstractText
Prostate MR spectroscopic imaging (MRSI) at 3T may provide two-fold higher spatial resolution over 1.5T, but this can result in longer acquisition times to cover the entire gland using conventional phase-encoding. In this study, flyback echo-planar readout trajectories were incorporated into a Malcolm Levitt's composite-pulse decoupling sequence (MLEV)-point-resolved spectroscopy sequence (PRESS) to accelerate the acquisition of large array (16 x 16 x 8), high spatial (0.154 cm(3)) resolution MRSI data by eight-fold to just 8.5 minutes. Artifact free, high-quality MRSI data was obtained in nine prostate cancer patients. Easy data reconstruction and the robustness of the flyback echo-planar encoding make this technique particularly suitable for the clinical setting. The short acquisition time provided by this method reduces the 3T prostate MRI/MRSI exam time, allows longer repetition times, and/or allows the acquisition of additional MR acquisitions within the same exam.
pubmed:grant
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
1053-1807
pubmed:author
pubmed:copyrightInfo
(c) 2007 Wiley-Liss, Inc.
pubmed:issnType
Print
pubmed:volume
25
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1288-92
pubmed:dateRevised
2007-12-3
pubmed:meshHeading
pubmed:year
2007
pubmed:articleTitle
High-speed 3T MR spectroscopic imaging of prostate with flyback echo-planar encoding.
pubmed:affiliation
Department of Radiology, University of California at San Francisco, San Francisco, CA 94143, USA.
pubmed:publicationType
Journal Article, Research Support, N.I.H., Extramural