2005-2-3
|
pubmed:abstractText |
Hyperpolarized (3)He MRI of the human lung was performed at 0.54 and 1.5 T using identical software and hardware (except for RF coils) at both field strengths. The T(*) (2) of (3)He gas in the lung was measured, and the effects of magnetic-susceptibility-induced field inhomogeneities on the appearance of interleaved-spiral and interleaved-echo-planar lung images at 1.5 T were compared to those at 0.54 T. Mean T(*) (2) values for (3)He gas in the healthy human lung were 26.8 +/- 1.5 ms and 67.9 +/- 1.3 ms at 1.5 and 0.54 T, respectively. At 0.54 T, interleaved-spiral images showed markedly less blurring due to susceptibility effects compared to images acquired at 1.5 T. At both 0.54 and 1.5 T, interleaved-echo-planar images appeared essentially identical to corresponding GRE images, even though the data-sampling period per echo and echo time were substantially longer for the interleaved-echo-planar images acquired at 0.54 T.
|
pubmed:language |
eng
|
pubmed:journal | |
pubmed:citationSubset |
IM
|
pubmed:chemical | |
pubmed:status |
MEDLINE
|
pubmed:month |
Jan
|
pubmed:issn |
0740-3194
|
pubmed:author | |
pubmed:copyrightInfo |
Copyright 2004 Wiley-Liss, Inc.
|
pubmed:issnType |
Print
|
pubmed:volume |
53
|
pubmed:owner |
NLM
|
pubmed:authorsComplete |
Y
|
pubmed:pagination |
212-6
|
pubmed:dateRevised |
2006-11-15
|
pubmed:meshHeading | |
pubmed:year |
2005
|
pubmed:articleTitle |
Hyperpolarized 3He lung imaging at 0.5 and 1.5 Tesla: a study of susceptibility-induced effects.
|