pubmed:abstractText
Hyperpolarized (3)He MRI of the human lung was performed at 0.54 and 1.5 T using identical software and hardware (except for RF coils) at both field strengths. The T(*) (2) of (3)He gas in the lung was measured, and the effects of magnetic-susceptibility-induced field inhomogeneities on the appearance of interleaved-spiral and interleaved-echo-planar lung images at 1.5 T were compared to those at 0.54 T. Mean T(*) (2) values for (3)He gas in the healthy human lung were 26.8 +/- 1.5 ms and 67.9 +/- 1.3 ms at 1.5 and 0.54 T, respectively. At 0.54 T, interleaved-spiral images showed markedly less blurring due to susceptibility effects compared to images acquired at 1.5 T. At both 0.54 and 1.5 T, interleaved-echo-planar images appeared essentially identical to corresponding GRE images, even though the data-sampling period per echo and echo time were substantially longer for the interleaved-echo-planar images acquired at 0.54 T.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jan
pubmed:issn
0740-3194
pubmed:author
pubmed:copyrightInfo
Copyright 2004 Wiley-Liss, Inc.
pubmed:issnType
Print
pubmed:volume
53
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
212-6
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2005
pubmed:articleTitle
Hyperpolarized 3He lung imaging at 0.5 and 1.5 Tesla: a study of susceptibility-induced effects.
pubmed:affiliation
Center for In-vivo Hyperpolarized Gas MR Imaging, Department of Radiology, University of Virginia School of Medicine, Charlottesville, VA 22908, USA.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't