Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
12
pubmed:dateCreated
2005-1-17
pubmed:abstractText
Monochromatic imaging can provide better contrast and resolution than conventional broadband radiography. In broadband systems, low energy photons do not contribute to the image, but are merely absorbed, while high energy photons produce scattering that degrades the image. By tuning to the optimal energy, one can eliminate undesirable lower and higher energies. Monochromatization is achieved by diffraction from a single crystal. A crystal oriented to diffract at a particular energy, in this case the characteristic line energy, diffracts only those photons within a narrow range of angles. The resultant beam from a divergent source is nearly parallel, but not very intense. To increase the intensity, collimation was performed with polycapillary x-ray optics, which can collect radiation from a divergent source and redirect it into a quasi parallel beam. Contrast and resolution measurements were performed with diffracting crystals with both high and low angular acceptance. Testing was first done at 8 keV with an intense copper rotating anode x-ray source, then 17.5 keV measurements were made with a low power molybdenum source. At 8 keV, subject contrast was a factor of five higher than for the polychromatic case. At 17.5 keV, monochromatic contrast was two times greater than the conventional polychromatic contrast. The subject contrasts measured at both energies were in good agreement with theory. An additional factor of two increase in contrast, for a total gain of four, is expected at 17.5 keV from the removal of scatter. Scatter might be simply removed using an air gap, which does not degrade resolution with a parallel beam.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0094-2405
pubmed:author
pubmed:issnType
Print
pubmed:volume
31
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
3288-97
pubmed:dateRevised
2008-11-21
pubmed:meshHeading
pubmed:year
2004
pubmed:articleTitle
Beam collimation with polycapillary x-ray optics for high contrast high resolution monochromatic imaging.
pubmed:affiliation
Health Imaging Division, Eastman Kodak Company, Rochester, New York 14650, USA.
pubmed:publicationType
Journal Article, Comparative Study, Research Support, U.S. Gov't, Non-P.H.S., Evaluation Studies