pubmed:abstractText
Measurement of vascular access flow (QVA) has been suggested as a method of choice for vascular access quality (VAQ) monitoring. Besides traditional duplex Doppler, a number of bedside methods based mostly on the Krivitski principle of QVA evaluation from recirculation at reversed needles (RX), have been developed. This work compares ultrasonic dilution (UD), taken as a reference, HD01, Transonic Systems; duplex Doppler (DD); thermodilution (TD), BTM, Fresenius; optodilutional RX measurement (ORX), Critline III, R-mode, HemaMetrics; direct optodilutional QVA evaluation from jumpwise changes in ultrafiltration rate at both normal and reversed needles connection (OABF), Critline III, ABF-mode; and direct transcutaneous optodilutional QVA evaluation (TQA), Critline III TQA. Firstly, reproducibility of each method was assessed by duplicate measurement at unchanged conditions. This was followed by paired measurement with each method performed at controlled change in relevant measurement condition (two different extracorporeal blood flows in UD and TD, changed sensor position in TQA). Finally paired measurements by each method and the reference method performed at identical conditions were evaluated to assess accuracy of each method. The simple Krivitski formula QVA=QB(1-RX)/RX was used wherever manual QVA calculation was needed. Very high reproducibility was seen in UD, both for measurement at the same extra corporeal blood flow (QB) (correlation coefficient of duplicate measurement r=0.9702, n=58) and for measurement at two different QB (r=0.9735, n=24), justifying its current status of a reference method in QVA evaluation. Slightly lower reproducibility of TD measurement at the same QB (r=0.9197, n=40) and at two different QB (r=0.8508, n=168) can be easily overcome by duplicate measurement with averaging. High correlation of TD vs. UD (r=0.9543, n=54) makes TD a viable clinical alternative in QVA evaluation. Consistently different QVA obtained at two different QB should prompt closer investigation of anatomical conditions of the access. Use of the simple Krivitski formula in TD (which measures total recirculation, i.e. sum of access recirculation and cardiopulmonary recirculation) brings about underestimation of QVA, which progressively increases from QVA of about 600 mL/min up. Good correlation, although with significant scatter (r=0.8691, n=27) was found between the DD- and UD-based QVA. By far the worst reproducibility at the same QB from among the investigated methods was found in ORX (0.6430, n 23). Also the correlation of ORX vs. UD was lower than in other methods (r=0.702, n=33) and general overestimation of QVA by about 25% was noted. Correlation of OABF vs. UD (r=0.6957, n=26) was slightly better than that of ORX and it gave less overestimated values. The TQA method showed very high reproducibility (r=0.9712, n=85), however only for unchanged sensor position. Correlation of QVA measured at two different sensor positions was much worse (r=0.7255, n=22). Correspondence of TQA vs. UD was satisfactory (r=0.8077, n=36). Skilled and experienced operators are a must with this method.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Dec
pubmed:issn
0391-3988
pubmed:author
pubmed:issnType
Print
pubmed:volume
26
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
1056-63
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2003
pubmed:articleTitle
Comparison of different techniques of hemodialysis vascular access flow evaluation.
pubmed:affiliation
General University Hospital, Department of Medicine, Prague-Strahov, Czech Republic. f.lopot@vfn.cz
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't, Evaluation Studies