pubmed:abstractText
Pronation and supination have been shown to affect wrist goniometer measurement accuracy. The purpose of this study was to compare differences in measurement accuracy between a commonly used biaxial, single transducer wrist goniometer (System A) and a biaxial, two-transducer wrist goniometer (System B) over a wide range of pronation and supination (P/S) positions. Eight subjects moved their wrist between -40 and 40 degrees of flexion/extension (F/E) and -10 and 20 degrees of radial/ulnar (R/U) deviation in four different P/S positions: 90 degrees pronation; 45 degrees pronation; 0 degrees neutral and 45 degrees supination. System A was prone to more R/U crosstalk than System B and the amount of crosstalk was dependent on the P/S position. F/E crosstalk was present with both goniometer systems and was also shown to be dependent on P/S. When moving from pronation to supination, both systems experienced a similar extension offset error; however R/U offset errors were roughly equal in magnitude but opposite in direction. The calibration position will affect wrist angle measurements and the magnitude and direction of measurement errors. To minimize offset errors, the goniometer systems should be calibrated in the P/S posture most likely to be encountered during measurement. Differences in goniometer design and application accounted for the performance differences.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Oct
pubmed:issn
1050-6411
pubmed:author
pubmed:issnType
Print
pubmed:volume
12
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
413-20
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2002
pubmed:articleTitle
Comparison of measurement accuracy between two wrist goniometer systems during pronation and supination.
pubmed:affiliation
Department of Environmental Health, University of Washington, Box 357234, Seattle, WA 98195, USA. petej@u.wahington.edu
pubmed:publicationType
Journal Article, Comparative Study, Research Support, Non-U.S. Gov't