Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
6
pubmed:dateCreated
2001-2-6
pubmed:abstractText
We present the results of a multicenter evaluation of a new point-of-care system (Cardiac Reader) for the quantitative determination of cardiac troponin T (CARDIAC T Quantitative test) and myoglobin (CARDIAC M test) in whole blood samples. The Cardiac Reader is a CCD camera that optically reads the immunochemical test strips. The measuring range is 0.1 to 3 microg/l for CARDIAC T Quantitative and 30 to 700 microg/l for CARDIAC M. Both tests are calibrated by the manufacturer. The reaction times of the tests are 12 or 8 minutes, respectively. Method comparisons were performed with 281 heparinized blood samples from patients with suspected acute coronary syndromes. The results obtained with CARDIAC T Quantitative showed a good agreement compared with cardiac troponin T ELISA (r = 0.89; y = 0.93x + 0.02). The method comparison between CARDIAC M and Tina-quant Myoglobin also showed a good agreement between both assays (r = 0.98; y = 0.92x + 1.6). Test lot-to-lot comparisons yielded differences of 2% and 6% for CARDIAC T Quantitative and of 0 to 11% for CARDIAC M. The within-run imprecision with blood samples and control materials was acceptable for CARDIAC T Quantitative (CV 10 to 15%) and good for CARDIAC M (CV 5 to 10%). The between-instrument CV was below 7% for CARDIACT Quantitative and below 5% for CARDIAC M. The cross-reactivity of CARDIAC T Quantitative with skeletal troponin T was approximately 0.003%. No significant analytical interference was detected for any of the assays in investigations with biotin (up to 100 microg/l), hemoglobin (up to 0.125 mmol/l), hematocrit (26 to 52%), bilirubin (up to 340 micromol/l), triglycerides (up to 5.0 mmol/l), and 18 standard drugs. With the Cardiac Reader reliable quantitative results can be easily obtained for both cardiac markers. The system is, therefore, particularly suitable for use in emergency rooms, coronary care units and small hospitals.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:chemical
pubmed:status
MEDLINE
pubmed:month
Jun
pubmed:issn
1434-6621
pubmed:author
pubmed:issnType
Print
pubmed:volume
38
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
567-74
pubmed:dateRevised
2006-11-15
pubmed:meshHeading
pubmed:year
2000
pubmed:articleTitle
Evaluation of a point-of-care system for quantitative determination of troponin T and myoglobin.
pubmed:affiliation
Medizinische Klinik II, Medizinische Universität zu Lübeck, Germany.
pubmed:publicationType
Journal Article, Research Support, Non-U.S. Gov't, Multicenter Study, Evaluation Studies