pubmed:abstractText
An important issue in intensity modulated radiation therapy (IMRT) is the verification of the monitor unit (MU) calculation of the planning system using an independent procedure. Because of the intensity modulation and the dynamic nature of the delivery process, the problem becomes much more involved than that in conventional radiation therapy. In this work, a closed formula for MU calculation is derived. The approach is independent of the specific form of leaf sequence algorithms. It is straightforward to implement the procedure using a simple computer program. The approach is illustrated by a simplified example and is demonstrated by a few CORVUS (NOMOS Corporation, Sewickley, PA) treatment plans. The results indicate that it is robust and suitable for IMRT MU verification.