pubmed:abstractText
We present a newly developed torsional stiffness measurement device with the potential to quantitatively assess the in vivo torsional stiffness of bone regenerate during distraction osteogenesis. We describe the form and function of this device and its application in a model of regenerate consolidation. The device was able to produce data to assess stiffness of the regenerate with an accuracy between +/- 3 and +/- 9% for material stiffness ranging between 0.1 and 2.4 Nm/o and with a precision of +/- 3.6%. This method provides advantages over similar methods of bone fracture healing assessment with guaranteed maintenance of bone axis, minimized risk of bone misalignment during the bone healing process and a close relation to the functional loading pattern in torsion of bones such as tibia and femora.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
IM
pubmed:status
MEDLINE
pubmed:month
Aug
pubmed:issn
0021-9290
pubmed:author
pubmed:issnType
Print
pubmed:volume
32
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
857-60
pubmed:dateRevised
2009-11-11
pubmed:meshHeading
pubmed:year
1999
pubmed:articleTitle
A new device to quantify regenerate torsional stiffness in distraction osteogenesis.
pubmed:affiliation
Department of Trauma Surgery, Virchow Klinikum, Humboldt-University, Berlin, Germany. windhagen@annastift.de
pubmed:publicationType
Journal Article