Statements in which the resource exists as a subject.
PredicateObject
rdf:type
lifeskim:mentions
pubmed:issue
1
pubmed:dateCreated
1991-5-7
pubmed:abstractText
The lacrimal drainage system drains fluid that exists normally on the surface of the eye into the nasal cavity. This paper describes a hand-held, personal-computer-based instrument to quantitatively estimate the resistance to fluid flow through the lacrimal drainage system. This resistance measurement is proposed as a possible means to diagnose blockage of the system and to verify the degree of change in resistance after clinical treatment. The instrument system consists of a hand-held syringe attached to pressure and flow transducers. Resistance is calculated as the ratio of differential pressure to flow rate and is displayed in real time on a computer monitor.
pubmed:language
eng
pubmed:journal
pubmed:citationSubset
H
pubmed:status
MEDLINE
pubmed:issn
0363-8855
pubmed:author
pubmed:issnType
Print
pubmed:volume
16
pubmed:owner
NLM
pubmed:authorsComplete
Y
pubmed:pagination
57-60
pubmed:dateRevised
2004-11-17
pubmed:meshHeading
pubmed:articleTitle
Instrument to test patency of lacrimal drainage system.
pubmed:affiliation
Mechanical and Aerospace Engineering, West Virginia University, Morgantown.
pubmed:publicationType
Journal Article